Reverse hip replacement system: addressing instability concerns or introducing new risks? A scoping review
- Authors: Khalifa A.A.1,2
-
Affiliations:
- South Valley University
- Aster Sanad Hospital
- Issue: Vol 32, No 2 (2026)
- Pages: 149-156
- Section: Reviews
- Submitted: 17.03.2026
- Accepted: 23.04.2026
- Published: 12.05.2026
- URL: https://journal.rniito.org/jour/article/view/17859
- DOI: https://doi.org/10.17816/2311-2905-17859
- ID: 17859
Cite item
Full Text
Abstract
Background. To address instability after total hip arthroplasty (THA), a Reverse Hip Replacement System (RHRS) has been proposed, inspired by the concept of reverse total shoulder arthroplasty.
The aim of the review — to gather reported results, explore early outcomes, and identify potential challenges of the reverse hip replacement system.
Methods. A systematic search in PubMed and Scopus databases was performed in March 2025 using predetermined search terms, which revealed an initial 656 results. Only four studies were eligible for inclusion: one biomechanical study and three clinical studies (one case series and two case reports).
Results. The three clinical studies included 24 patients aged 60 to 79 years; 13 (54.2%) were male, and all were diagnosed with advanced hip osteoarthritis. After follow-up ranging from 3 weeks to 24 months, all patients were satisfied with the results, except for 2 (8.3%). Detailed function and PROMs outcomes for 22 patients at 24 months of follow-up showed significant improvement in WOMAC, HHS, OHS, HOOS, SF-36 physical component summary (PCS), and ED-5D (p < 0.05). However, there was a nonsignificant change in the SF-36 mental component summary (MCS) (p = 0.272). Radiological outcomes, as determined by RSA analysis, at 24 months for 18 acetabular cups showed a mean migration of 0.087±0.152 mm. The overall complication rate was 29.2% 7 (cases): one intraoperative calcar crack, one deep periprosthetic joint infection, one postoperative traumatic dislocation, one femoral neuropraxia, one occasional pain, and two recalcitrant iliotibial band syndrome. The biomechanical study results showed lower polyethylene wear debris equivalent to other systems, no wear debris due to metal-on-metal contact, and no signs of polyethylene edge or backside wear.
Conclusion. Besides the promising biomechanical results, early functional and radiological outcomes of the reverse hip replacement system are encouraging; however, the incidence of complications is considerably high. Moreover, the results should be interpreted cautiously owing to the small number of included patients and the short follow-up period.
Full Text
INTRODUCTION
There is no doubt that total hip arthroplasty (THA) is one of the most successful surgical procedures in orthopedics. Nevertheless, surgeons, in collaboration with bioengineers and other relevant parties, are continually seeking improvements to the THA prosthesis to minimize its potential drawbacks and enhance its longevity [1, 2].
One crucial complication, which is a leading cause of revision after THA, is instability or dislocation [3, 4, 5]. A. Upfill-Brown et al. re-ported instability as the most common reason for THA revision in the US between 2012 and 2018, accounting for 20.4% of revision causes [6]. Furthermore, a recent study by L.L. Hermansen et al., based on evaluating 5,415 primary THAs from the Danish Hip Arthroplasty register, reported a 1-year dislocation rate of 2.8% (CI 2.4-3.3), with 37% of these patients having recurrent dislocations [7].
To avoid or minimize THA instability, in addition to surgical approaches and patient selection, various solutions related to implant design and materials have been introduced, such as large heads, dual-mobility articulations, and constrained liners [3]. Moreover, owing to the relative similarity between the hip and shoulder joints and inspired by the successful outcomes of the reverse total shoulder arthroplasty (RTSA) [8, 9], the concept of Reverse Hip Replacement System (RHRS) or reverse THA was proposed. In this design, the cobalt-chrome ball is located on a trunion within the acetabular cup rather than the femoral stem and articulates with a polyethylene liner attached to a femoral cup1.
This device was recently approved for compassionate use by the Food and Drug Administration (FDA), but the clinical results remain insufficient. The scoping review performed in the present study aims this knowledge gap and determine whether the theoretical stability advan-tages of RHRS outweigh the added complexity of two tapers, new wear interfaces, and untested long-term behavior.
The aim of the review — to gather reported results, explore early outcomes, and identify potential challenges of the reverse hip replacement system.
METHODS
Search strategy and selection criteria
A systematic search was conducted in March 2025, and this scoping review followed the Preferred Reporting Items for Systematic Reviews and Meta-Analyses Extension for Scoping Reviews (PRISMA-ScR) guidelines to identify studies related to reverse THA [10].
The inclusion criteria were any article reporting on RHRS clinical results (functional, radiological, and complications) or biomechanical testing results. Studies that lacked the desired outcomes were excluded.
A comprehensive literature search was conducted using two databases (PubMed and Scopus) with the following search terms: ((total hip arthroplasty [Title/Abstract]) OR (total hip replacement [Title/Abstract])) AND (reverse [Title/Abstract]).
No restrictions were applied to the study type or language, and the search was conducted from the database's inception through March 2025. The resultant studies were downloaded into EndNote 20, where duplicates were identified and removed. An initial screening for eligibility was then carried out based on the title and abstract.
The full texts of eligible studies were evaluated for inclusion, and the reference lists of included articles were searched for additional eligible studies. The initial search yielded 656 articles eligible for initial assessment, of which only four were ultimately included. These were identified from the databases and a subsequent reference search, comprising three clinical studies [11, 12, 13] and one biomechanical study [14]. Details on the number of citations identified and on what was included or excluded at each screening step, along with the justification for exclusions, are presented according to the PRISMA flowchart (Figure 1).
Figure 1. Flow diagram of article search and selection
Data extraction and critical appraisal
The following information was extracted from included studies: author(s); date of publication, country of ori-gin, study type; the final number of patients; follow-up period, surgical details (including surgical approach, prosthesis fixation, and so on), clinical outcomes, radiological outcomes, and complications. For the biomechanical study, the testing details and variables evaluated were reported. Furthermore, any relevant outcomes reported by any of the studies were collected.
RESULTS
Clinical studies
The data from the clinical trials are systematized and presented in Table 1.
Table 1. Basic demographic characteristics, clinical, radiological, and complications outcomes reported in the included clinical studies
Characteristics | A.V. Jr. Lombardi, J.B. Adams [11] | T.R. Turgeon et al. [12] | S.J. Zabinski [13] | |||
Year | 2024 | 2023 | 2023 | |||
Country | USA | Canada | USA | |||
Study type | Case report | Case series | Case report | |||
No. of patients | 1 | 23 initially, and only 22 were finally analyzed | 1 | |||
Gender | M | 11 M, 11 F | M | |||
Age, years | 60 | 70.6±3.5 (65 to 79) | 64 | |||
Diagnosis | OA of the left hip | OA | OA of the right hip | |||
BMI, kg/m2 | 33.9 | 31.0±5.7 | NR | |||
Follow-up | Three weeks | 24 months | Six months | |||
Surgical approach | Minimally invasive direct lateral approach | Posterior or lateral (per surgeon preference, as four surgeons were involved) | Anterior approach | |||
Anesthesia | NR | Spinal anesthesia with sedation | NR | |||
Prothesis details | Hemispherical press-fit acetabular shell:
|
| NR | |||
Fixation | Cementless fixation, 3 acetabular screws | Cementless fixation, acetabular screws in all cases | Cementless fixation, 2 acetabular screws | |||
Rehabilitation protocol | NR | Weight-bearing as tolerated and no hip precautions for all patients | No changes | |||
Radiological outcomes | At three weeks: AP and lateral X-rays showed well-fixed components in satisfactory position and alignment | RSA at 24 months showed the following:
| NR | |||
Functional outcomes | pain-free, tolerating physical therapy, and an acceptable range of motion and surgical outcomes | Score | Preoperative | Last FU | p | Feeling well, and showed excellent motion, full strength, and no instability |
WOMAC | 56.3±2.5 | 10.0±13.3 | < 0.001 | |||
HHS | 50.5±4.6 | 88.9±11.8 | < 0.001 | |||
OHS | 18.0±6.6 | 42.4±7.4 | < 0.001 | |||
HOOS | 37.1±13.3 | 87.4±15.9 | < 0.001 | |||
SF-36 PCS | 29.9±8.8 | 46.1±11.7 | < 0.001 | |||
SF-36 MCS | 52.9±13.2 | 56.0±8.4 | 0.272 | |||
ED-5D | 0.740±0.099 | 0.847±0.159 | 0.004 | |||
Complications | None |
| Occasional pain | |||
OA — osteoarthritis; NR — not reported; RSA — radiostereometric analysis; FU — follow-up; PJI — periprosthetic joint infection; DAIR — debridement, antibiotics, and implant retention.
Basic characteristics
The three clinical studies included were one case series from Canada [12], and two case reports from the US [11, 13], both of which were part of an ongoing clinical trial (ClinicalTrials.gov ID NCT05357378). The three studies involved a total of 25 patients aged 60 to 79 years (one patient was excluded from the final analysis [12]). All patients (13 males and 11 females)were diagnosed with end-stage hip osteoarthritis. The surgery was performed through a minimally invasive direct lateral approach in one patient [11], a direct anterior approach in one patient [13], and a lateral or posterior approach (per surgeon preference) in 23 patients [12].
All studies used the same implant (HIT Hip Replacement System, Hip Innovation Technology, USA), and all were cementless fixation. On the acetabular side, supplementary screws were placed in all cups (two screws in 14 patients, three screws in nine patients, and one screw in one patient).
Clinical outcomes
Regarding patient satisfaction, 22 (91.7%) patients were satisfied with the procedure and outcomes; however, two (8.3%) patients [12] reported being somewhat dissatisfied.
Detailed functional and PROMs outcomes were only reported for 22 patients by T.R. Turgeon et al. [12], where after a mean follow-up of two years, the authors reported significant improvement in Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Harris Hip Score (HHS), Oxford Hip Score (OHS), Hip disability and Osteoarthritis Outcome Score (HOOS), 36-item Short Form survey v. 2 physical component summary (SF-36 PCS), and EuroQol five-dimension health questionnaire (ED-5D) (p < 0.05). However, they reported a nonsignificant change in the SF-36 mental component summary (MCS) (p = 0.272). One patient still reported occasional pain at the six-month follow up [13].
Radiological outcomes
A.V. Jr. Lombardi and J.B. Adams reported that the X-rays showed well-positioned implants at three weeks without further details [11]. T.R. Turgeon et al. reported the radiological results using radiostereometric analysis (RSA) to assess com-ponent fixation and migration [12]. At two years of follow-up, the mean acetabular component migration by 24 months was 0.087±0.152 mm, while the mean femoral component migration was 0.002±0.194 mm. The authors reported that the values were significantly lower than the threshold reported in the literature (0.2 mm for the acetabular component, p = 0.005; and 0.5 mm for the femoral component, p < 0.001).
Complications
Seven (29.2%) complication incidents were reported as follows: one (4.2%) deep periprosthetic joint infection (PJI) required DAIR, then two-stage exchange; one (4.2%) patient experienced femoral neuropraxia (did not require intervention and resolved during the follow-up period); one (4.2%) patient had intraoperative calcar crack (managed with a single cerclage cable); two (8.4%) patients had recalcitrant iliotibial band syndrome (managed by local injections); traumatic dislocation in one (4.2%) patient which occurred on postoperative day 4 (required closed reduction under anesthesia); and occasional pain till the last follow-up was reported in one (4.2%) patient.
Biomechanical study
L. Braddon et al. evaluated the corrosion at the tapers (acetabular and femoral) and polyethylene wear characteristics [14].
Taper's evaluation characteristics
Visual evaluation of the acetabular and femoral taper surfaces after 50 million cycles revealed no evidence of surface damage (gouges, scratches, or structural damage).
Grading the corrosion levels, using a five-grade scale
After 10 million cycles, all tapers showed a score of 2 or 3 (indicating discoloration of the taper with minimal pitting or etch marks) after evaluating their superior, anterior, inferior, and posterior faces. After 50 million cycles of femoral taper only, it scored 2 to 4 (indicating between discoloration to some level of pitting and etch marks).
Morse taper connections disassembly forces assessment
Femoral taper: after 10 and 50 million cycles of dynamic fatigue, it was 12.957±1797 and 12.506±74 N, respectively. Acetabular taper: after 10 and 50 million cycles of dynamic fatigue, it was 3429±347 and 4329±443 N, respectively.
Polyethylene wear characteristics
This was evaluated while the acetabular cup placement inclination (ranging from 20° to 70°)/anteversion (ranging from 0° to 40°). While there was no gait separation for 5 million cycles, the results showed a mean wear rate of 1.60±0.45 mg/mc at 45°/0°, 1.30 mg/mc at 20°/0°, 1.71 mg/mc at 70°/0°, 1.02 mg/mc at 20°/40°, and 1.27 mg/mc at 70°/40°. With a gait separation of 2 mm for 1 million cycles, it showed a mean wear rate of 7.03±0.39 mg/mc at 45°/0°, 6.31 mg/mc at 20°/0°, 7.08 mg/mc at 70°/0°, 5.71 mg/mc at 20°/40°, and 5.76 mg/mc at 70°/40°. No edge wear was reported at any of the tested positions.
DISCUSSION
Although the results published on the newly proposed RHRS are scarce and considered immature, they seem promising, with acceptable functional and radiological outcomes. The complication rate might be considered high; however, most complications were transient and did not require surgical intervention. Furthermore, biomechanical testing demonstrated acceptable prosthesis behavior even under extreme implant positioning conditions.
According to L. Braddon et al., the RHRS is composed of two tapers, one to connect the acetabular ball (cobalt-chrome) to the acetabular shell, while the other connects the femoral cup (highly cross-linked high molecular weight polyethylene) with the femoral stem [14]. The claim primarily relied on providing a more stable design for older and more active populations, with the potential to improve longevity. Although most of the included patients showed acceptable functional outcomes, up to a 91% satisfaction rate, and none of the included patients required revision for implant-related issues, these claims need to be proven based on long-term clinical data, especially with the limited follow-up provided in the included studies.
Regarding radiological outcomes, T.R. Turgeon et al. reported implant migration measured by RSA using tantalum beads inserted before implant placement [12]. They noted that at 12 and 24 months, the reported migration of the acetabular and femoral components was below the reported limits in the literature, leading to a > 5% revision risk at 10 years [15]. However, the authors did not comment on the target zone for acetabular shell positioning, unlike A.V. Jr. Lombardi and J.B. Adams, who reported aiming at the classic acetabular orientation of 45° of abduction and 20° of anteversion [11].
It is worth noting that in the biomechanical study by L. Braddon et al., the authors evaluated the prosthesis wear characteristics under extreme acetabular shell positioning up to 70° of abduction and 40° of anteversion, with no increase in the edge wear [14].
Although the reported complication incidents were considerably high (29.2%), most of them were treated nonsurgically, except for one deep PJI case that required revision. T.R. Turgeon et al. reported one incident of traumatic dislocation after the patient had a fall. The authors treated the dislocation by closed reduction under sedation. They reported that the patient was satisfied with the outcome at the last follow-up, with no other incidents of instability [12].
Lessons learned from previous catastrophic failures with newly introduced hip implants should alert surgeons to the challenges and concerns that may be associated with newly introduced implants, such as RHRS [16, 17, 18]. A.V. Jr. Lombardi and J.B. Adams reported that by reversing the implant geometry, as in the RTSA designs, the contact area between the acetabular and femoral components will expand, enhancing stability at the extremes of motion with minimal instability risk [11]. However, the anatomical and biomechanical variations between the hip and shoulder joints should be considered when arguing that one implant design is applicable to the other [19, 20].
Furthermore, adding two tapers might theo-retically double the risk of trunion-related issues such as taper corrosion. Although T.R. Turgeon et al. reported no problems related to adverse reactions to metal debris [12], two years of follow-up is too early to conclude complete safety [21].
Lastly, one of the proposed advantages of RHRS was maintaining the load forces perpendicular to the femoral polyethylene cup throughout the flexion range of motion, assuming this would result in prosthesis stability less dependent on the individual components' positioning [12, 22, 23]. However, none of the authors defined the target for implant positioning. Furthermore, if the implant was designed to address instability issues, it is expected to be used for patients with spino-pelvic issues. Future studies should determine whether specific target zones or an insertion algorithm can be defined for RHRS implantation.
Study limitations
The study has several inherent limitations. First, the small number of included studies and the correspondingly small number of included patients. Second, the very short follow-up period limits the evaluation of longevity and possible long-term complications. Third, all the included patients were diagnosed with hip osteoarthritis, which carries no high risk for dislocation, to challenge the proposed higher stability of RHRS. Fourth, a potential conflict of interest arises from the manufacturing company's financial support for all included studies. Fifth, there was only one biomechanical study for evaluating the RHRS, which relied mainly on corrosion and polyethene wear evaluation, where further evaluation could be performed, such as liner push-out and lever-out testing, femoral stem fatigue loading in a worst-case configuration, micromotion measurements at the bone-implant interface, impingement-free range of motion, and jump distance. Sixth, although concerns have arisen over the past decade regarding the use of fixed target safe zone numbers for acetabular cup placement, one study reported placing acetabular cups within those classic numbers. Lastly, only two databases were searched; furthermore, owing to the small number of included studies and their nature, detailed quantitative analysis of the results was not possible. Therefore, the current review should be considered a descriptive mapping exercise, and it was not possible to estimate true event rates or to make comparative claims versus conventional THA.
CONCLUSION
Although the reverse hip replacement system showed promising early radiological and functional outcomes, with relatively fewer complications requiring surgical intervention or revision, the biomechanical assessment revealed wear rates and behavior comparable to those of the respective conventional hip implants. The results should be cautiously interpreted owing to the selected patient category, only those with osteoarthritis, with no higher risks for instability, and the short clinical follow-up. Therefore, drawing solid, promising conclusions was not possible until longer follow-up studies reporting the in vivo wear and safety profile are conducted.
DISCLAIMERS
Funding source. This study was not supported by any external sources of funding.
Disclosure competing interests. The author declare that they have no competing interests.
Ethics approval. Not applicable.
Consent for publication. Not required.
Use of artificial intelligence. No generative artificial intelligence technologies were used in the preparation of this manuscript.
ДОПОЛНИТЕЛЬНАЯ ИНФОРМАЦИЯ
Источник финансирования. Автор заявляет об отсутствии внешнего финансирования при проведении исследования.
Возможный конфликт интересов. Автор декларирует отсутствие явных и потенциальных конфликтов интересов, связанных с публикацией настоящей статьи.
Этическая экспертиза. Не применима.
Информированное согласие на публикацию. Не требуется.
Генеративный искусственный интеллект. При создании статьи технологии генеративного искусственного интеллекта не использовали.
1 Hip Innovation Technology L. Reverse HRS Product Overview. 2015. Available from: http://www.hipinnovationtechnology.com/product-information-innovation.html.
About the authors
Ahmed A. Khalifa
South Valley University; Aster Sanad Hospital
Author for correspondence.
Email: ahmed_adel0391@med.svu.edu.eg
ORCID iD: 0000-0002-0710-6487
MD, FRCS, MSc, Assistant Professor, Qena faculty of medicine and University Hospital
Egypt, Qena; Saudi Arabia, RiyadhReferences
- Fontalis A., Epinette J.A., Thaler M., Zagra L., Khanduja V., Haddad F.S. Advances and innovations in total hip arthroplasty. SICOT J. 2021;7:26. doi: 10.1051/sicotj/2021025.
- Li Z. Advancements of biomaterial in hip replacement technology incorporating ceramic materials. J Orthop. 2024;62:27-35. doi: 10.1016/j.jor.2024.09.021.
- Takahashi D., Shimizu T., Miyazaki T., Ogawa T., Iwasaki N. Instability in total hip arthroplasty: A literature review. J Joint Surg Res. 2023;(1):92-96. doi: 10.1016/j.jjoisr.2023.03.004.
- Sirignano M.N., Nessler J.M., Rhea E.B., Ong K.L., Watson H.N., Yakkanti M.R. et al. Incidence of Instability Following Primary Total Hip Arthroplasty Continues to Decline in the Medicare Population. J Arthroplasty. 2023;38(7S):S89-S94.e1. doi: 10.1016/j.arth.2023.04.035.
- Hernigou P., Barbier O., Chenaie P. Hip arthroplasty dislocation risk calculator: evaluation of one million primary implants and twenty-five thousand dislocations with deep learning artificial intelligence in a systematic review of reviews. Int Orthop. 2023; 47(2):557-571. doi: 10.1007/s00264-022-05644-2.
- Upfill-Brown A., Hsiue P.P., Sekimura T., Patel J.N., Adamson M., Stavrakis A.I. Instability Is the Most Common Indication for Revision Hip Arthroplasty in the United States: National Trends From 2012 to 2018. Arthroplast Today. 2021;11:88-101. doi: 10.1016/j.artd.2021.08.001.
- Hermansen L.L., Iversen T.F., Iversen P., Viberg B., Overgaard S. The "true" 1-year incidence of dislocation after primary total hip arthroplasty: validation of an algorithm identifying dislocations in the Danish National Patient Register based on 5,415 patients from the Danish Hip Arthroplasty Register. Acta Orthop. 2024;95:380-385. doi: 10.2340/17453674.2024.41064.
- Zeng G.J., Sheng X., Lie D.T.T. Short- to medium-term outcomes and future direction of reverse shoulder arthroplasty: Current concepts. J ISAKOS. 2023;8(6): 398-403. doi: 10.1016/j.jisako.2023.10.005.
- Franceschi F., Giovannetti de Sanctis E., Gupta A., Athwal G.S., Di Giacomo G. Reverse shoulder arthroplasty: State-of-the-art. J ISAKOS. 2023;8(5): 306-317. doi: 10.1016/j.jisako.2023.05.007.
- Tricco A.C., Lillie E., Zarin W., O'Brien K.K., Colquhoun H., Levac D. et al. PRISMA Extension for Scoping Reviews (PRISMA-ScR): Checklist and Explanation. Ann Intern Med. 2018;169(7):467-473. doi: 10.7326/M18-0850.
- Lombardi A.V. Jr., Adams J.B. Use of a Novel Reverse Hip Replacement System to Address Dislocation and Instability. Surg Technol Int. 2024;44:263-270. doi: 10.52198/24.STI.44.OS1798.
- Turgeon T.R., Hedden D.R., Bohm E.R., Burnell C.D. Radiostereometric analysis and clinical outcomes of a novel reverse total hip system at two years. Bone Jt Open. 2023;4(5):385-392. doi: 10.1302/2633-1462.45.BJO-2023-0018.R1.
- Zabinski S.J. Case Study: Reverse Hip Replacement System May Provide Greater Stability — Especially with Spinal Pelvic Disorders. OrthoSpineNews. 2023. Available from: https://orthospinenews.com/2023/ 12/27/case-study-reverse-hip-replacement-system-may-provide-greater-stability-especially-with-spinal-pelvic-disorders/.
- Braddon L., Termanini Z., MacDonald S., Parvizi J., Lieberman J., Frankel V. et al. Corrosion and Tribology of Materials Used in a Novel Reverse Hip Replacement. Materials (Basel). 2017;10(7):751. doi: 10.3390/ma10070751.
- van der Voort P., Pijls B.G., Nieuwenhuijse M.J., Jasper J., Fiocco M., Plevier J.W. et al. Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies. Acta Orthop. 2015;86(5):575-585. doi: 10.3109/17453674.2015.1043832.
- Ozan F., Kahraman M., Baktır A., Gençer K. Catastrophic failure and metallosis of the acetabular component in total hip arthroplasty. J Orthop Surg Res. 2021;16(1): 349. doi: 10.1186/s13018-021-02492-5.
- Barakat N., Browne J.A., Cui Q. Catastrophic Early Failure and Fragmentation of a Modern Moderately Cross-linked Polyethylene Acetabular Liner. Arthroplast Today. 2023;22:101161. doi: 10.1016/j.artd.2023.101161.
- Khalifa A.A., Bakr H.M. Updates in biomaterials of bearing surfaces in total hip arthroplasty. Arthroplasty. 2021;3(1):32. doi: 10.1186/s42836-021-00092-6.
- Jaffar A.A., Abass S.J., Ismael M.Q. Biomechanical Aspects of Shoulder and Hip Articulations: a Comparison of Two Ball and Socket Joints. Al-Khwarizmi Engineering J. 2006;2(1):1-14.
- Fabrocini B., Mercaldo N. A Comparison between the Rotator Cuffs of the Shoulder and Hip. Strength and Conditioning J. 2003;25(4):63-68. doi: 10.1519/00126548-200308000-00017.
- Eltit F., Wang Q., Wang R. Mechanisms of Adverse Local Tissue Reactions to Hip Implants. Front Bioeng Biotechnol. 2019;7:176. doi: 10.3389/fbioe.2019.00176.
- Crowninshield R.D., Maloney W.J., Wentz D.H., Humphrey S.M., Blanchard C.R. Biomechanics of large femoral heads: what they do and don't do. Clin Orthop Relat Res. 2004;(429):102-107.
- Berliner J.L., Regalado-Magdos A., Ma C.B., Feeley B.T. Biomechanics of reverse total shoulder arthro-plasty. J Shoulder Elbow Surg. 2015;24(1):150-160. doi: 10.1016/j.jse.2014.08.003.
Supplementary files